5.2 Check and adjustment of the datum point
[Check]
Set up the instrument on the tripod halfway between two walls or
pillars about 40m away so that the ‘X side’of the instrument faces the
‘Wall Aand level it.
Operate the instrument and turn on the power to the detector.(use the
detection accuracy initially set.)
Mark the reference point detected for both walls.(A,B)
Arrange a pair of target plates (with the same scale) and fix them to the
walls (A and B) so that the center values of the target plates coincide
with the reference point marks.
Loosen the center screw of the tripod to turn the instrument 180 for
the ‘X side’to face the ‘Wall B’and tighten the center lock screw again
for the instrument to be leveled again.
1
2
3
4
5
Over view
(Keypad)
X side